Geoffrey Lentner edited figures/xrf_energy_calibration2/caption.tex  about 8 years ago

Commit id: 91461b7be88989bbe6a48f3c0e6696ddcd9f5da7

deletions | additions      

       

Replace this text *panel-bottom:* All the calibration spectra (Ba, Cu, Rb, Tb) are stacked for display purposes. The blue, downward pointing triangular markers indicate the emission peaks that were fit\* for the calibration analysis. All partner pairs (e.g. K$\alpha_1$ and K$\alpha_2$) were fitted as blended, double gaussians over a linear background (an 8 parameter $X^2$ minimization).  *panel-middle:* All peak centroids from the bottom panel are represented  with your caption the currently reported energies from the literature. The horizontal error bars are scaled to the FWHM of the fitted gaussians for each corresponding emission peak. The uncertainties reported for each transition energy are too small to be represented meaningfully here. A linear regression gives the reported calibration curve. The results and statistics for this regression are provided in the figure.  *panel -top:* The standard residuals to the linear regression (middle panel) are provided.  \* These corresponding to the $K\alpha_1$, $K\alpha_2$, $K\beta_1$, and $K\beta_2$. Other features exist in these spectra but were ignored for the purpose of the calibration analysis.
  [NIST X-ray Transitions Energy Database](http://physics.nist.gov/PhysRefData/XrayTrans/Html/search.html)