\caption{\label{tab:standardctr} Standard coincidence apparatus measurements for two identical polished Proteus LYSO:Ce scintillator crystals wrapped in PTFE tape. All crystals have a cross section of $2\times2$ mm$^2$. $\chi^2_\text{Reduced}$ is the reduced chi-squared fit to the delay peak.}
\begin{tabular}{cccccccc}
\hline
Wrapped By & length (mm) & Left Energy Resolution (\%) & Right Energy Resolution (\%) & Detected $\gamma\gamma$ Events &
Valid $\gamma\gamma$ Events & Delay Peak Centroid (ps) & CTR (ps)\\
\hline
Wrap 1 & 5 & 7.05$\pm$0.07 & 9.44$\pm$0.09 & 856$\pm$29 & 11177$\pm$106 & -68.8$\pm$2.4 & 154.9$\pm$4.9 \\
Wrap 2 & 5 & 6.33$\pm$0.08 & 7.36$\pm$0.09 & 361$\pm$19 & 6344$\pm$80 & -68.8$\pm$3.4 & 140.8$\pm$6.7 \\