\begin{table} \caption{\label{tab:referencevals} Coincidence time resolution values for two identical polished $2\times2\times5$mm$^3$ Ca-co-doped LSO:Ce wrapping in PTFE tape for standard and DOI measurements. $\chi_\text{Reduced}$ is the reduced chi-squared fit to the delay peak distribution.} \begin{tabular}{ccccccc} %\begin{tabular}{ccccccc}  \begin{tabular}{cp{1cm}p{2.3cm}p{2.3cm}p{1.7cm}p{2.2cm}c}  Coincidence Apparatus & Left Energy Resolution (\%) & Right Energy Resolution (\%) & Detected $\gamma\gamma$ Events & Delay Peak Centroid (ps) & $\sigma_\textrm{ref}$ (ps) & CTR (ps)\\ \hline Standard & 7.8$\pm$0.1 & 9.8$\pm$0.1 & 637$\pm$25 & -77.5$\pm$2.3 & 39.3$\pm$1.2 & 131.0$\pm$4.0 \\ DOI & 10.5$\pm$0.1 & 12.7$\pm$0.1 & 1498$\pm$39 & -14.0$\pm$1.5 & 39.6$\pm$0.9 & 132.0$\pm$2.9 \\ \hline \end{tabular} \end{table}