Mark S. Brown edited Standard table.tex  over 10 years ago

Commit id: bfb0ea28547a98bdfb3228192da286270cf8432b

deletions | additions      

       

\begin{table}  \caption{\label{tab:standardctr} Standard coincidence apparatus measurements for two identical polished Proteus LYSO:Ce scintillator crystals wrapping in PTFE tape. All crystals have a cross section of $2\times2$mm$^2$. $\chi_\text{Reduced}$ $\chi^2_\text{Reduced}$  is the reduced chi-squared fit to the delay peak.} \begin{tabular}{lrlllllr}  configuration & length (mm) & Left Energy Resolution (\%) & Right Energy Resolution (\%) & Detected $\gamma\gamma$ Events & Delay Peak Centroid (ps) & CTR (ps) & $\chi^2_\text{Reduced}$ \\  \hline  wrapped & 5 & 7.05$\pm$0.07 & 9.44$\pm$0.09 & 856$\pm$29 & -68.8$\pm$2.4 & 154.9$\pm$4.9 & 1.2 \\ wrapped & 10 & 6.66$\pm$0.05 & 7.59$\pm$0.05 & 515$\pm$23 & -168.4$\pm$3.6 & 169.7$\pm$7.0 & 0.9 \\ wrapped & 10 & 8.24$\pm$0.06 & 9.63$\pm$0.07 & 1032$\pm$32 & -124.4$\pm$2.6 & 185.6$\pm$5.0 & 0.8 \\ wrapped & 15 & 6.82$\pm$0.04 & 7.67$\pm$0.04 & 1569$\pm$40 & -108.8$\pm$2.0 & 178.2$\pm$3.6 & 1.0 \\ wrapped & 15 & 9.05$\pm$0.05 & 9.07$\pm$0.05 & 1875$\pm$43 & -176.0$\pm$2.1 & 201.9$\pm$4.1 & 0.8 \\ wrapped & 20 & 10.97$\pm$0.05 & 9.85$\pm$0.05 & 1731$\pm$42 & -85.2$\pm$2.2 & 202.7$\pm$4.0 & 1.4 \\ wrapped & 30 & 9.73$\pm$0.04 & 10.79$\pm$0.05 & 3502$\pm$59 & -70.6$\pm$1.6 & 212.4$\pm$3.0 & 1.4 \\ wrapped & 30 & 10.94$\pm$0.04 & 13.07$\pm$0.06 & 3382$\pm$58 & -58.6$\pm$1.8 & 237.7$\pm$3.3 & 2.1 \\  \hline  \end{tabular}  \end{table}