Mark S. Brown deleted other30mmresults.tex  over 10 years ago

Commit id: 677ecce4b5d19f91d06735fd18d37831583e1a2d

deletions | additions      

       

Appendix-Method.tex  figures/Run_23972vs24044Reference_00005_CTR/Run_23972vs24044Reference_00005_CTR.svg  figures/Run_23972vs24044Reference_00005_IndexData/Run_23972vs24044Reference_00005_IndexData.svg  other30mmresults.tex           

\subsubsection{Other 30mm Results}   In figure \ref{fig:centroidposition} the shift in the delay peak position with increasing DOI is plotted. It can be seen that a plateau is reached for both configurations at 20mm. In \cite{Moses_Derenzo_1999} this plateau is attributed to the travel time between photons travelling 'towards' and 'away' from the photodetector approaching equality. As the electronic trigger from the discriminator is tuned as close to the beginning of the signal as possible this cannot come before generated light has physically travelled from it's emission position to the photodetector. This is supported by the total shift seen of 150ps to 370ps corresponding to an approximate physical distance of 30mm.