Mark S. Brown edited Reference Measurements.tex
over 10 years ago
Commit id: 47628cdb094093107698bdbf3e41e6308cbf662b
deletions | additions
diff --git a/Reference Measurements.tex b/Reference Measurements.tex
index d738ab3..3b15779 100644
--- a/Reference Measurements.tex
+++ b/Reference Measurements.tex
...
\section{Reference
Detector Measurements}
The reference scintillator crystal, shown on the left of figure \ref{fig:actualsetup}, is a 2×2×5mm3 Ca-co-doped LSO:Ce wrapping in PTFE tape coupled. Using two identical such crystals the coincidence time resolution was determined using both the standard and DOI coincidence apparatus. The measured values are shown in table \ref{tab:referencevals}. The standard coincidence CTR value is in agreement with prior measurement\cite{arron_Meyer_Pauwels_Lecoq_2012}. Unless otherwise stated all values are subtracted in quadrature using the reference value of $131\pm4$ps.