A solenoid with a coil constant of 15mT/A was used to apply a magnetic field to polarize the water. The magnetic field of the solenoid, which was the polarizing field, could be varied using a variable current input. The field of the Helmholtz coils, which artificially increased or decreased the Earth's magnetic field applied to the sample, was adjusted using a separate variable current input. The gradient coils can be used to change the gradient of the magnetic field in the x, y, and z directions. In our experiment, the gradient coils were used to create a homogeneous field but were otherwise not adjusted over the course of the experiment. \\\\
A bandpass amplifier, part of the NMR instrument, was tuned to resonate at the Larmor precession frequency. The bandpass amplifier restricted the frequencies to those around the Larmor precession frequency, and reduced any background noise such as 60 Hz fluctuations in the field due to fluorescent lighting in the room, as well as amplified the precession frequency.
\\\\ \\