Error analysis(Johnson Noise)  The discrepancy between our result ($1.46 \times 10^-23 \text{m2 kg s-2 K-1}$) and the accepted Boltzmann constant ($1.38 \times 10^-23 \text{m2 kg s-2 K-1}$)is approximately 5.8\% . Both sets of data we obtained (with a $1K\Omega$ $1\textrm{K\Omega}$  resister and a $10K\Omega$) $10\textrm{K\Omega}$)  gave a consistent value for the Boltzmann constant which is a matter of accuracy rather than precision, so this error is most likely a systematic error. Assuming that our thermometer is accurate (for room temperature), the temperature within the instrument, where the resister is, may be higher. As we can see in Eq.~ \ref{eq:boltzmann}, with a higher $\textrm{T}$, we would have a lower $\textrm{K}_\textrm{B}$ value which is opposite to what our value is, so this is unlikely what is causing the error.