Madeline Horn edited section_Dertermining_the_Unknown_Sample__.tex  over 8 years ago

Commit id: 3d45ca7fc6fbff977533cd319f29402e7c7887a4

deletions | additions      

       

In order to determine what element the unknown sample was, we kept all the same settings from Trials 1 and 2, and kept the calibrations the same in order to find the peak values of the unknown sample. We placed the unknown sample into the scintillator at the fifth slot from the top and the sample was within an old film container. The sample ran for $3670$ seconds Live Time. The difference between Live Time and Real Time is that as the program is producing the spectrum, there is a slight lag as it processes the data. So, the Live Time is the actual time the data is being recorded. By keeping the calibrations from trials 1 and 2, we were able to run the sample and have the data in Energy versus intensity instead of bins versus intensity. Beause we know the energy of the peaks it made it simple to find the element of the unknown sample.  From the Smith College Physics radiation safety protocols version 5.0, we know that the only radioactive materials the sample could be are Cesium-137, Cobalt 60, Sodium 22, and Strontium 90. This reduced the error in our unknown sample because that left us with only four possible elements. As you can see in FIGURE WHAT, there are two peaks. After fitting gaussians to each peak (using Igor Pro), the first peak is at $0.53655 \pm 7.11 \cdot 10^(-5) 10^{-5}  \textrm{ MeV}$ and the second is at $1.2791 \pm 0.000406 \textrm{ MeV}$