Will edited Our_beam_had_a_power__.tex  over 8 years ago

Commit id: d49884a21761c624e97570b5ad839305b404df4b

deletions | additions      

       

Our beam had a power of 1 mW. It passed through the rubidium cell, which contained both rubidium 85 and rubidium 87, and hit a photodiode. The photodiode was connected to an SRS fast amplifier with a gain of 1250, which was connected to an oscilloscope from which we could collect data. We also logged data using To monitor the frequency of our light, we used  a \textbf{Bristol Wavemeter}. Out attempt to use the wavemeter as a frequency calibration ultimately failed. The intent was to calibrate the time axis of our transmission graph with a frequency. Unfortunately the wavemeter was not stable enough across a single scan to do this. This stability contributes an additional systematic error that was too time intensive to study for this experiment.  We scanned the laser over a frequency range for each of the electronic  transitions, tuning the central frequency of our scan  to make sure that the absorption was happening in the middle of each scan so as to get undistorted data. scan.  The absorption curve for Doppler spectroscopy is a convolution between a Gaussian and a Lorentzian, but can be we  fit to a Gaussian. Gaussian for the reasons described in section