Our beam had a power of 1 mW. It passed through the rubidium cell, which contained both rubidium 85 and rubidium 87, and hit a photodiode. The photodiode was connected to an SRS fast amplifier with a gain of 1250, which was connected to an oscilloscope from which we could collect data.
We also logged data using To monitor the frequency of our light, we used a \textbf{Bristol Wavemeter}.
Out attempt to use the wavemeter as a frequency calibration ultimately failed. The intent was to calibrate the time axis of our transmission graph with a frequency. Unfortunately the wavemeter was not stable enough across a single scan to do this. This stability contributes an additional systematic error that was too time intensive to study for this experiment.
We scanned the laser over a frequency range for each of the
electronic transitions, tuning
the central frequency of our scan to make sure that the absorption was happening in the middle of each
scan so as to get undistorted data. scan. The absorption curve for Doppler spectroscopy is a convolution between a Gaussian and a Lorentzian, but
can be we fit to a
Gaussian. Gaussian for the reasons described in section