this is for holding javascript data
Klaus edited Results2.tex
over 8 years ago
Commit id: f4cb82f273f901611292f8a0f23b283b61db1b07
deletions | additions
diff --git a/Results2.tex b/Results2.tex
index a0612c3..cf254ac 100644
--- a/Results2.tex
+++ b/Results2.tex
...
ThunderSTORM's option for Multi-emitter Fitting Analysis (MFA) produces excellent results with recognising and separating overlapping EBCCD photon events, as shown in Fig~\ref{fig_mfa}. For the USAF data containing an average of 150 photons/frame the photon count increases 2\% with MFA enabled.
Centroiding was then applied to higher photon density
biological cell fluorescence microscopy data
containing of a fixed cell labelled with xxxx. The frames contain an average of 460 photons/frame (Fig~\ref{fig_cell}). With 7-pixel PSF fitting radius and MFA enabled, the photon count increases 13.0\% compared to using 2-pixel fitting radius, and 5.2\% compared to 7-pixel fitting radius without MFA. Fitting of multiple PSFs is time-consuming and the processing time increases significantly with MFA enabled, as expected (see Table~\ref{table:results}). However, for biological imaging the image acquisition time is a critical parameter, and the separation of overlapping events can potentially lead to significant reduction in image acquisition times.