Handheld ED-XRF spectrometers in geochemical investigation - the comparative studies for glacial deposits (Spitsbergen)
  • +4
  • Lidia Kozak,
  • Juliana Silva Souza,
  • Adam Nawrot,
  • Jedrzej Proch,
  • Marcin Kazmierski,
  • Agnieszka Zawieja,
  • Przemyslaw Niedzielski
Lidia Kozak
Adam Mickiewicz University, Adam Mickiewicz University, Adam Mickiewicz University
Author Profile
Juliana Silva Souza
Adam Mickiewicz University, Adam Mickiewicz University, Adam Mickiewicz University
Author Profile
Adam Nawrot
Institute of Geophysics Polish Academy of Sciences, Institute of Geophysics Polish Academy of Sciences, Institute of Geophysics Polish Academy of Sciences
Author Profile
Jedrzej Proch
Adam Mickiewicz University, Adam Mickiewicz University, Adam Mickiewicz University
Author Profile
Marcin Kazmierski
Adam Micvkiewicz Universsity, Adam Micvkiewicz Universsity, Adam Micvkiewicz Universsity
Author Profile
Agnieszka Zawieja
MEWO, MEWO, MEWO
Author Profile
Przemyslaw Niedzielski
Adam Mickiewicz University in Poznan Poland, Adam Mickiewicz University in Poznan Poland, Adam Mickiewicz University in Poznan Poland

Corresponding Author:[email protected]

Author Profile

Abstract

This study presents the determination of the content of selected metals: Ba, Ca, Fe, Nb, Rb, Sr, Y, Zn, Zr in postglacial deposits from two glacial valleys (Ebbadalen and Elsadalen) in the Petunia Bay (southern Spitsbergen). Deposits analyses were performed using X-ray fluorescence (XRF) in parallel with two portable spectrometers from different manufacturers to investigate the accuracy and reliability of the instruments. Statistical analysis of the results indicated that the measurements carried out with two spectrometers are statistically significantly different, which is probably due to the different calibration characteristics used by the manufacturers of XRF spectrometers. However, the analysis of the spatial distribution of element concentrations using Geographic Information System (GIS) tools showed that the distribution maps of element concentrations were similar regardless of the spectrometer used in the analyses.