The accuracy of heartbeat detection using photoplethysmography technology in cardiac patients
  • +2
  • Sebastiaan Blok,
  • Martijn Piek,
  • Igor Tulevski,
  • Aernout Somsen,
  • Michiel Winter
Sebastiaan Blok
Cardiologie Centra Nederland

Corresponding Author:[email protected]

Author Profile
Martijn Piek
Cardiologie Centra Nederland
Author Profile
Igor Tulevski
Cardiologie Centra Nederland
Author Profile
Aernout Somsen
Cardiologie Centra Nederland
Author Profile
Michiel Winter
Cardiologie Centra Nederland
Author Profile

Abstract

Introduction Photoplethysmography (PPG) in wearable sensors potentially plays an important role in accessible heart rhythm monitoring. We investigated the accuracy of a state-of-the-art bracelet (Corsano 287) for heartbeat detection in cardiac patients and evaluated the efficacy of a signal qualifier in identifying medically useful signals. Methods Patients from an outpatient cardiology clinic underwent a simultaneous resting ECG and PPG recording, which we compared to determine accuracy of the PPG sensor for detecting heartbeats within 100 and 50 ms of the ECG-detected heart beats and correlation and Limits of Agreement for heartrate (HR) and RR-intervals. We defined subgroups for skin type, hair density, age, BMI and gender and applied a previously described signal qualifier. Results In 180 patients 7,914 ECG-, and 7,880 (99%) PPG-heartbeats were recorded. The PPG-accuracy within 100 ms was 94.6% (95%CI 94.1-95.1) and 89.2% (95%CI 88.5-89.9) within 50 ms. Correlation was high for HR (R = 0.991 (95%CI 0.988-0.993), n = 180) and RR-intervals (R = 0.891 (95%CI 0.886-0.895), n = 7880). The 95% Limits of Agreement (LoA) were -3.89 to 3.77 (mean bias 0.06) beats per minute for HR and -173 to 171 (mean bias -1) for RR-intervals. Results were comparable across all subgroups. The signal qualifier led to a higher accuracy in a 100 ms range (98.2% (95%CI 97.9-98.5)) (n=143). Conclusion We showed that the Corsano 287 Bracelet with PPG-technology can determine HR and RR-intervals with high accuracy in cardiovascular at-risk patient population among different subgroups, especially with a signal quality indicator.